IME SA provides a wide range of liquid analyzers using different analysis modules according to the characteristics of the liquids, the components you intend to analyse and the applications that they are designed for.
The Fluoride Analyzer measures the fluoride in drinking water applications in order to control the Fluoride that is added or that comes in excess from the underground.
This Analyzer measures the fluoride content in potable water, proving to be extremely effective and accurate up to 100 mg/L F -.
MEASURING PRINCIPLE:
- Sample comparison by the measuring cell which then observes the electrochemical reaction in a selective electrode of fluoride ions.
COMMON APPLICATIONS:
- Monitoring of fluoride levels of spring water;
- Monitoring drinking water fluoridation;
- Monitoring the defluorination process.
KEY FEATURES:
- Windows based menu options display allows user-friendly and intuitive use;
- Presentation of results from graphs with automatic calculation of statistical trends;
- Communication flexibility through Ethernet connectivity, 6 isolated current outputs, 10 configurable relay alarms, Profibus DP V1.0 and SD Memory slot;
- Option of simple or varied sample collection (up to a maximum of 3) with sample collection containers integrated and alarms in case of sample collection failure;
- Simplified fluid handling the simple piston pump withdraws a precise volume of sample and reagent through a manifold valve into the optical measuring cell and the movement of the piston;
- Advanced optical system with the LED light source intensity automatically adjusted at calibration to eliminate drift and compensate for any cell fouling.
MEASURING RANGE:
MEASURING PERFORMANCE:
- Accuracy: ≤ ±2% of reading or ± 0,02ppm (whichever is the greater);
- Repeatability: ≤ ± 2% of reading or ± 0,02ppm (whichever is greater);
- Resolution: 0 to 10ppm – 0,001ppm, 10 to 100ppm - 0,01ppm;
- Calibration: 2-point automatic calibration, with the option of manual initiation. The interval between automatic calibrations manually selectable from four times a day to one a week.